Distortion
Generate a sinusoidal test signal with 1 V offset and 0.5 V amplitude. Use 100 Hz, 1 kHz, 10 kHz, 100 kHz and 500 kHz test signal. Perform this test with WSLX-880-002-M-H14-T-PD load, where the current monitor is used for the current measurement. Also use a 1 ohm load and measure the load voltage, which, in this case, is linear to the current. Use the internal waveform generator as well as an external one connected to the analog setpoint. Calculate THD; include FFT images created from raw data - do not use oscilloscope screenshots